User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: KeyTek ESD-1:ESD Test System DATE: 04-Aug-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 6 NUMBER OF LINES: 91 CONFIGURATION: Fluke 5700A CONFIGURATION: HP 34401A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 IEEE [@34401]*RST 1.004 IEEE [@34401]FUNC "VOLT:DC"[GTL] 1.005 HEAD INITIAL CONDITIONS 1.006 DISP Connect the input to the DCA-1 attenuator to 1.006 DISP the HI/LO OUTPUT from the 5700. 1.007 DISP Connect the DCA-1 output to the HI/LO INPUT on the 1.007 DISP 34401. 1.008 5700 1000V S 2W 1.009 IEEE [@34401]READ?[I][GTL] 1.010 5700 * S 1.011 MATH m[1] = (mem1 / mem) / 1000 1.012 DISP Disconnect the DCA-1 from the 5700. 1.013 HEAD {} 1.014 HEAD {* High Voltage Tests *} 1.015 HEAD {DN-1 Network} 1.016 HEAD High Voltage Tests 1.017 JMP 2.001 1.018 EVAL 2.001 DISP On the PSC-1 press the red OFF button. 2.002 DISP Insert the DN-1 discharge network into the ESD-1 gun. 2.003 DISP Set the gun polarity to +. 2.004 DISP Connect the DCA-1 to the DN-1 network. 2.005 MATH m[20] = 0 2.006 CALL KeyTek High Voltage Sub 2.007 HEAD {} 2.008 HEAD {DN-2 Network} 2.009 HEAD High Voltage Tests 2.010 JMP 3.001 2.011 EVAL 3.001 DISP Disconnect the DCA-1 from the DN-1 network. 3.002 DISP Disconnect the DN-1 network from the ESD-1 gun. 3.003 DISP Connect the DN-2 network to the ESD-1 gun. 3.004 DISP Connect the DCA-1 to the DN-2 network. 3.005 MATH m[20] = 0 3.006 CALL KeyTek High Voltage Sub 3.007 HEAD {} 3.008 HEAD {DN-10 Network} 3.009 HEAD High Voltage Tests 3.010 JMP 4.001 3.011 EVAL 4.001 DISP On the PSC-1 press the red OFF button. 4.001 DISP Disconnect the DCA-1 from the DN-2 network. 4.002 DISP Disconnect the DN-2 network from the ESD-1 gun. 4.003 DISP Connect the DN-10 network to the ESD-1 gun. 4.004 DISP Connect the DCA-1 to the DN-10 network. 4.005 MATH m[20] = 0 4.006 CALL KeyTek High Voltage Sub 4.007 HEAD {} 4.008 HEAD {DN-X S/N 8906288 Network} 4.009 HEAD High Voltage Tests 4.010 JMP 5.001 4.011 EVAL 5.001 DISP On the PSC-1 press the red OFF button. 5.001 DISP Disconnect the DCA-1 from the DN-10 network. 5.002 DISP Disconnect the DN-10 network from the ESD-1 gun. 5.003 DISP Connect the DN-X S/N 8906288 network to the ESD-1 gun. 5.004 DISP Connect the DCA-1 to the DN-X network. 5.005 MATH m[20] = 0 5.006 CALL KeyTek High Voltage Sub 5.007 HEAD {} 5.008 HEAD {DN-X S/N 8906286 Network} 5.009 HEAD High Voltage Tests 5.010 JMP 6.001 5.011 EVAL 6.001 DISP On the PSC-1 press the red OFF button. 6.001 DISP Disconnect the DCA-1 from the DN-X network. 6.002 DISP Disconnect the DN-X network from the ESD-1 gun. 6.003 DISP Connect the DN-X S/N 8906286 network to the ESD-1 gun. 6.004 DISP Connect the DCA-1 to the DN-X network. 6.005 MATH m[20] = 1 6.006 CALL KeyTek High Voltage Sub 6.007 DISP Disconnect the DCA-1 from the 34401. 6.008 DISP Remove the DCA-1 and the DN-X network from the ESD-1. 6.009 IEEE [@34401]*RST[GTL] 6.010 HEAD {} 6.011 END